Analytical Chemistry Core Lab Inorganics

TOCTN-L

The Shimadzu TOC-L analyzers is an essential tool in environmental science and can be used to analyze a variety of carbon species total carbon (TC), organic carbon (TOC), non-purgable organic carbon (NPOC), inorganic carbon (IC), dissolved organic carbon (DOC) and dissolved inorganic carbon (DIC) in liquid or solid samples using catalytic oxidation. Additional module provides total nitrogen (TN) analysis in water samples.
Sample: Drinking, waste, and seawater, soil, and sediment. Minimum of 50 ml for liquid and 5 g for solid.

Shimadzu
TOC-L-CSH
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Simultaneous Thermal Analyzer with QMS (STA449 F1 with QMS 403C

Simultaneous Thermal Analysis generally refers to the simultaneous application of Thermogravimetry (TGA) and Differential Scanning Calorimetry (DSC) to one and the same sample in one instrument. The test conditions are perfectly identical for the TGA and DSC signals (same atmosphere, gas flow rate, vapor pressure on the sample, heating rate, thermal contact to the sample crucible and sensor, radiation effect, etc.). Furthermore, it improves sample throughput as more information is gathered from each test run. It also has a QMS Mass spectrometer for further identification of evolved gases.

Netzsch
STA 449 F1 Jupiter
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Simultaneous Thermal Analyzer with Humidity Generator

Simultaneous Thermal Analysis generally refers to the simultaneous application of Thermogravimetry (TGA) and Differential Scanning Calorimetry (DSC) to one and the same sample in one instrument. The test conditions are perfectly identical for the TGA and DSC signals (same atmosphere, gas flow rate, vapor pressure on the sample, heating rate, thermal contact to the sample crucible and sensor, radiation effect, etc.). Furthermore, it improves sample throughput as more information is gathered from each test run. STA instrument can be also coupled with Humidity Generator. In this case, the humidified gas produced in the generator is transferred to the STA via a heated transfer line. The humidified gas can be used from 2 % to 98 % relative humidity. Measurements under conditions of controlled relative humidity provide information that is crucial for understanding the effects that moisture content can have on the properties of a wide range of materials. Users can also simultaneously measure sorption enthalpies and analyze the sorption results.

Netzsch
STA 449 F3 Jupiter
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Laser Flash Apparatus (LFA)

Laser Flash Diffusivity Testing Apparatus (LFA 447)
The flash technique is one of the most widely used methods for the determination of the thermal diffusivity and thermal conductivity of solids all over the world. Using this method, the front side of a plane-parallel sample with a well defined thickness is heated by a short light or laser pulse. The resulting temperature rise on the back surface is measured versus time using an infrared detector. Analyzing the measured detector signal with appropriate mathematical models yields information on thermal diffusivity and the specific heat of a material. Together with the density of the material the thermal conductivity can be determined. Additionally, analysis of multi-layer systems or the measurement of liquids is possible by the flash method.

NETZSCH
LFA 447/NanoFlash
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Karl Fisher

The Karl Fisher titrator uses coulometric titration to determine trace amounts of water in a sample, from ppm to 5%. A stromboli oven is available for samples that are unable to be directly added into the titration vessel.
Sample - petroleum, gels, semi-solids/creamy. Minimum sample amount is 5 mL or 5 g.

Mettler Toledo
C30
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Organic Elemental Analyzer-2

Determination of C,H,N,O content of sample in given matrix determination of total carbon, total organic carbon, hydrogen, nitrogen, sulphur and oxygen.
Sample - solids, oil (heavy oil, viscous) or sediments, with a minimum quantity of 10 mg.

Thermo Fisher
Flash 2000
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Organic Elemental Analyzer-1

Determination of C,H,N,O content of sample in given matrix determination of total carbon, total organic carbon, hydrogen, nitrogen, sulphur and oxygen.
Sample - solids, oil (heavy oil, viscous) or sediments, with a minimum quantity of 10 mg.

Thermo Fisher
Flash 2000
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Mill grinder

Planetary Ball Mills are used wherever the highest degree of fineness is required. Apart from the classical mixing and size reduction processes, the mills also meet all the technical requirements for colloidal grinding and have the energy input necessary for mechanical alloying processes. The extremely high centrifugal forces of a planetary ball mill result in very high pulverization energy and therefore short grinding times.

Retsch
PM 200
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

ICP-MS-3 Agilent 8900

Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry which is capable of detecting metals at low concentrations

Agilent Technologies
G3665 A
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

ICP-MS/MS (8800)

Detection of metals at low concentration using inductively coupled plasma tandem mass spectrometry (ICP-MS/MS) technique.

Agilent Technologies
G3663 A
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Viscometer_KEM

Viscosity is a critical characteristic of petroleum samples, bio-diesels, polymers, ionic liquids to evaluate the resistance of fluids to flow or stress. The instrument is capable of handling high temperature and at elevated pressure of polymers, petrochemicals, and many different challenging matrices in less than a minute run. This instrument will reduce the amount of sample to use and waste. The sample tube module is sealed to protect the users from hazardous sample.

Kyoto Electronics
EMS 1000
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

ICP-OES PE Optima

Quantitative and qualitative determination of metals in given matrix. Sample and calibration curve preparations are required prior to analysis.
Sample - minimum quantity of 5 mL for liquids and 250 mg for solids. Liquid sample needs to be acidified (1 % w/w, Nitric Acid) and transported in plastic container.

Perkin Elmer
Optima 8300
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

ICP-OES Agilent 5110

Quantitative and qualitative determination of metals in given matrix. Sample and calibration curve preparations are required prior to analysis.
Sample - minimum quantity of 5 mL for liquids and 250 mg for solids. Liquid sample needs to be acidified (1 % w/w, Nitric Acid) and transported in plastic container.

Agilent Technologies
G8015A
Building 3, Level 3 East
Analytical Chemistry Core Lab Metabolomics

GC-MS/MS Pesticide Analyzer-2

This is a gas chromatography-tandem mass spectrometry system that is utilized to determine polycyclic aromatic hydrocarbons (PAHs) and a GC-amenable pesticides, among other (semi-)volatile organic compounds.

Agilent Technologies
G7012B
Building 3, Level 3 East
Analytical Chemistry Core Lab Organics

GC-MS/MS Pesticide Analyzer-1

This is a gas chromatography-tandem mass spectrometry system that is utilized to determine polycyclic aromatic hydrocarbons (PAHs) and a GC-amenable pesticides, among other (semi-)volatile organic compounds.

Agilent Technologies
G7012A
Building 3, Level 3 East
Analytical Chemistry Core Lab Inorganics

Fusion

The Eagon 2 instrument is completely automatic. It has two independent fusion positions that can work synchronously or asynchronously to increase productivity.

Panalytical
Eagon 2/943004700001
Building 3, Level 3 East
Analytical Chemistry Core Lab Organics

GPC-Dual Angle Light Scattering (DMF)

Gel Permeation Chromatograph is a classic tool for the measurement of average molecular weights and polydispersity index of polymers that are soluble at room temperature using a combination of refractive index and dual angle light scattering detectors.

Agilent Technologies
G1329B
Building 3, Level 3 East
Analytical Chemistry Core Lab Organics

GPC-RID (DMF)

Gel Permeation Chromatography is a classic tool for the measurement of average molecular weights and polydispersity index of polymers. This instrument is
particularly used for dimethylformamide-soluble polymers.

Agilent Technologies
G1312B
Building 3, Level 3 East
Analytical Chemistry Core Lab Organics

GPC-RID (Aqua)

Gel Permeation Chromatograph is a classic tool for the measurement of average molecular weights and polydispersity index of polymers. This instrument is
particularly used for water-soluble polymers.

Agilent Technologies
G1314D
Building 3, Level 3 East
Analytical Chemistry Core Lab Organics

MALDI-TOF-MS

Matrix-assisted laser desorption ionization-time of flight mass spectrometer is a widely used technique for the molecular weight measurement of synthetic molecules and polymers. It is especially useful for the determination of the monomeric units, repeating unit numbers, and end groups in polymer samples.

Bruker
GT0264G201
Building 3, Level 3 East